Related Experiment Video
Updated: Nov 3, 2025

Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Balloon-Expandable Valve Geometry After Transcatheter Aortic Valve Replacement in Low-Risk Patients With Bicuspid
Giorgio A Medranda1, Toby Rogers2, Brian J Forrestal1
1Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
Background:
Prospective bicuspid low-risk transcatheter aortic valve replacement (TAVR) registries' data demonstrated encouraging short-term results. Detailed data on transcatheter heart valve (THV) geometry after deployment using contemporary devices are lacking. This study sought to examine valve geometry after TAVR in patients with bicuspid aortic stenosis (AS).
Methods:
The study population was patients from the LRT (Low Risk TAVR) trial who underwent TAVR using the SAPIEN 3 THV for bicuspid and tricuspid AS. THV geometry measured on 30-day computed tomography (CT) included valve height, angle, depth, and eccentricity. Additionally, THV hemodynamics and outcomes post-TAVR were compared among patients with bicuspid and tricuspid AS.
Results:
A total of 107 patients from the LRT trial using the SAPIEN 3 THV were included in our analysis. On 30-day CT, the valve height ratio (1.07 vs. 1.07; p = 0.348), depths (right [5.6 mm vs. 6.2 mm; p = 0.223], left [5.3 mm vs. 4.4 mm; p = 0.082] and non [4.8 mm vs. 4.5 mm; p = 0.589] coronary cusps), eccentricities (1.08 vs. 1.07; p = 0.9550), and angles (except the right [3.9 degrees vs. 6.3 degrees; p = 0.003] and left [3.6 degrees vs. 6.0 degrees; p = 0.007]) were similar between bicuspid and tricuspid patients. Hemodynamics, stroke, and mortality were similar at 1 year.
Conclusion:
Despite challenging bicuspid anatomy of the aortic valve, our comprehensive CT analysis supports similar THV geometry between patients with bicuspid and tricuspid AS undergoing TAVR using the SAPIEN 3 THV in low-risk patients. This translated to excellent short-term clinical outcomes and THV hemodynamics in both aortic valve morphologies.
Trial Registry:
NCT02628899, https://clinicaltrials.gov/ct2/show/NCT02628899.
More Related Videos
12:17Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Mitral Stenosis III: Medical Management
Aortic Regurgitation III: Medical Management