MRI Highly Accelerated Wave-CAIPI T1-SPACE versus Standard T1-SPACE to detect brain gadolinium-enhancing lesions at
Augusto Lio M Goncalves Filho1,2, M Gabriela Figueiro Longo1, John Conklin1
1Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts, USA.
Background And Purpose:
High-resolution three-dimensional (3D) post-contrast imaging of the brain is essential for comprehensive evaluation of inflammatory, neoplastic, and neurovascular diseases of the brain. 3D T1-weighted spin-echo-based sequences offer increased sensitivity for the detection of enhancing lesions but are relatively prolonged examinations. We evaluated whether a highly accelerated Wave-controlled aliasing in parallel imaging (Wave-CAIPI) post-contrast 3D T1-sampling perfection with application-optimized contrasts using different flip angle evolutions (T1-SPACE) sequence (Wave-T1-SPACE) was noninferior to the standard high-resolution 3D T1-SPACE sequence for visualizing enhancing lesions with comparable diagnostic quality.
Methods:
One hundred and three consecutive patients were prospectively evaluated with a standard post-contrast 3D T1-SPACE sequence (acquisition time [TA] = 4 min 19 s) and an optimized Wave-CAIPI 3D T1-SPACE sequence (TA = 1 min 40 s) that was nearly three times faster than the standard sequence. Two blinded neuroradiologists performed a head-to-head comparison to evaluate the visualization of enhancing pathology, perception of artifacts, and overall diagnostic quality. A 15% margin was used to test whether post-contrast Wave-T1-SPACE was noninferior to standard T1-SPACE.
Results:
Wave-T1-SPACE was noninferior to standard T1-SPACE for delineating parenchymal and meningeal enhancing pathology (p < 0.01). Wave-T1-SPACE showed marginally higher background noise compared to the standard sequence and was noninferior in the overall diagnostic quality (p = 0.03).
Conclusions:
Our findings show that Wave-T1-SPACE was noninferior to standard T1-SPACE for visualization of enhancing pathology and overall diagnostic quality with a three-fold reduction in acquisition time compared to the standard sequence. Wave-T1-SPACE may be used to accelerate 3D post-contrast T1-weighted spin-echo imaging without loss of clinically important information.
More Related Videos
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
09:59A Magnetic Resonance Imaging Protocol for Stroke Onset Time Estimation in Permanent Cerebral Ischemia
Published on: September 16, 2017
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...