Establishing a reference focal plane using beads for trypan-blue-based viability measurements
Adele Peskin1, Steven P Lund2, Laura Pierce2
1National Institute of Standards and Technology, Boulder, Colorado.
Control beads ensure consistent image quality for trypan blue cell viability assays. This method improves accuracy and repeatability in automated cell analysis, overcoming challenges with cell-based autofocusing.
Area of Science:
- Biomedical Imaging
- Cell Biology
- Image Analysis
Background:
- Trypan blue dye exclusion assays for cell viability are sensitive to image quality and consistency.
- Automated image analysis requires reliable image acquisition at a consistent focal plane and adequate signal-to-noise ratio.
- Existing autofocusing methods struggle with cell sample heterogeneity, impacting accuracy and repeatability.
Purpose of the Study:
- To develop a robust method for ensuring consistent image acquisition in trypan blue assays.
- To improve the accuracy and repeatability of automated cell viability measurements.
- To introduce a novel autofocusing strategy using control beads.
Main Methods:
- Incorporation of control beads into cell samples for imaging.
- Utilizing bead image features with stable profiles across varying focal and exposure levels.
- Developing a predictive model based on bead image quality features to establish a reference focal plane.
- Implementing an open-source Graphical User Interface (BB-FAIT) for semi-automated analysis.
Main Results:
- A bead-based autofocusing method was developed, achieving over 95% accuracy.
- The method reproducibly returns to a reference focal plane using minimal beads (less than 3 per image).
- The approach demonstrates stability across a wide range of focal values and exposure levels.
- The Bead Benchmarking-Focus And Intensity Tool (BB-FAIT) was created to implement the method.
Conclusions:
- Control beads offer a reliable solution for maintaining consistent image quality in trypan blue assays.
- This bead-based approach enhances the accuracy and repeatability of automated cell viability measurements.
- The BB-FAIT tool provides a practical implementation for semi-automated cell analyzers.