Related Experiment Video
Updated: Nov 2, 2025

Quantification of Levator Ani Hiatus Enlargement by Magnetic Resonance Imaging in Males and Females with Pelvic Organ Prolapse
Published on: April 17, 2019
Manometric parameters, when measured with the 3-dimensional high-definition anorectal manometry probe, poorly predict
Matthew Woo1, Armaan Pandey1, Harman Gill2
1Division of Gastroenterology and Hepatology, University of Calgary, Calgary, AB, Canada.
Background:
There are little data evaluating the performance of the 3-dimensional high-definition anorectal manometry (3D-HDAM) system in the diagnosis of dyssynergic defecation. Physical properties of the thicker, rigid, 3D-HDAM probe may have implications on the measurements of anorectal pressures.
Aim:
Our aim was to compare 3D-HDAM to balloon expulsion test and magnetic resonance (MR) defecography.
Methods:
Consecutive constipated patients referred for anorectal function testing at the Calgary Gut Motility Centre (Calgary, Canada) between 2014 and 2019 were assessed. All patients underwent anorectal manometry with the 3D-HDAM probe, and a subset underwent BET or MR defecography. Anorectal manometric variables were compared between patients who had normal and abnormal BET.
Results:
Over the study period, 81 patients underwent both 3D-HDAM and BET for symptoms of constipation. 52 patients expelled the balloon within 3 minutes. Patients with abnormal BET had significantly lower rectoanal pressure differential (RAPD) (-61 vs. -31 mmHg for normal BET, p = 0.03) and defecation index (0.29 vs. 0.56, p = 0.03). On logistic regression analysis, RAPD (OR: 0.99, 95% CI: 0.97-0.99, p = 0.03) remained a negative predictor of abnormal BET. On ROC analysis, RAPD had an AUC of 0.65. There was good agreement between dyssynergic patterns on 3D-HDAM and defecographic evidence of dyssynergia (sensitivity 80%, specificity 90%, PLR 9, NLR 0.22, accuracy 85%).
Conclusions:
Manometric parameters, when measured with the 3D-HDAM probe, poorly predict prolonged balloon expulsion time. RAPD remains the best predictor of prolonged balloon expulsion time. The 3D-HDAM probe may not be the ideal tool to diagnose functional defecatory disorders.