Related Experiment Video
Updated: Nov 2, 2025

Longitudinal In Vivo Imaging and Quantification of Human Pancreatic Islet Grafting and Contributing Host Cells in the Anterior Eye Chamber
Published on: June 11, 2020
Deep learning image reconstruction algorithm for pancreatic protocol dual-energy computed tomography: image quality
Yoshifumi Noda1, Nobuyuki Kawai2, Shoma Nagata2
1Department of Radiology, Gifu University, 1-1 Yanagido, Gifu, 501-1194, Japan. noda1031@gifu-u.ac.jp.
Objectives:
To evaluate the image quality and iodine concentration (IC) measurements in pancreatic protocol dual-energy computed tomography (DECT) reconstructed using deep learning image reconstruction (DLIR) and compare them with those of images reconstructed using hybrid iterative reconstruction (IR).
Methods:
The local institutional review board approved this prospective study. Written informed consent was obtained from all participants. Thirty consecutive participants with pancreatic cancer (PC) underwent pancreatic protocol DECT for initial evaluation. DECT data were reconstructed at 70 keV using 40% adaptive statistical iterative reconstruction-Veo (hybrid-IR) and DLIR at medium and high levels (DLIR-M and DLIR-H, respectively). The diagnostic acceptability and conspicuity of PC were qualitatively assessed using a 5-point scale. IC values of the abdominal aorta, pancreas, PC, liver, and portal vein; standard deviation (SD); and coefficient of variation (CV) were calculated. Qualitative and quantitative parameters were compared between the hybrid-IR, DLIR-M, and DLIR-H groups.
Results:
The diagnostic acceptability and conspicuity of PC were significantly better in the DLIR-M group compared with those in the other groups (p < .001-.001). The IC values of the anatomical structures were almost comparable between the three groups (p = .001-.9). The SD of IC values was significantly lower in the DLIR-H group (p < .001) and resulted in the lowest CV (p < .001-.002) compared with those in the hybrid-IR and DLIR-M groups.
Conclusions:
DLIR could significantly improve image quality and reduce the variability of IC values than could hybrid-IR.
Key Points:
Image quality and conspicuity of pancreatic cancer were the best in DLIR-M. DLIR significantly reduced background noise and improved SNR and CNR. The variability of iodine concentration was reduced in DLIR.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...