Improvements in Glycemic Control Achieved by Altering the tmax Setting in the iLet® Bionic Pancreas When Using
Steven J Russell1,2, Courtney Balliro3, Magnus Ekelund4
1Diabetes Research Center, Massachusetts General Hospital, Boston, MA, USA. SJRUSSELL@mgh.harvard.edu.
Introduction:
We investigated the safety of, and glucose control by, the insulin-only configuration of the iLet® bionic pancreas delivering fast-acting insulin aspart (faster aspart), using the same insulin-dosing algorithm but different time to maximal serum drug concentration (tmax) settings, in adults with type 1 diabetes.
Methods:
We performed a single-center, single-blinded, crossover (two 7-day treatment periods) escalation trial over three sequential cohorts. Participants from each cohort were randomized to a default tmax setting (t65 [tmax = 65 min]) followed by a non-default tmax setting (t50 [tmax = 50 min; cohort 1], t40 [tmax = 40 min; cohort 2], t30 [tmax = 30 min; cohort 3]), or vice versa, all with faster aspart. Each cohort randomized eight new participants if escalation-stopping criteria were not met in the previous cohort.
Results:
Overall, 24 participants were randomized into three cohorts. Two participants discontinued treatment, one due to reported 'low blood glucose' during the first treatment period of cohort 3 (t30). Mean time in low sensor glucose (< 54 mg/dl, primary endpoint) was < 1.0% for all tmax settings. Mean sensor glucose in cohorts 1 and 2 was significantly lower at non-default versus default tmax settings, with comparable insulin dosing. The mean time sensor glucose was in range (70-180 mg/dl) was > 70% for all cohorts, except the default tmax setting in cohort 1. No severe hypoglycemic episodes were reported. Furthermore, there were no clinically significant differences in adverse events between the groups.
Conclusion:
There were no safety concerns with faster aspart in the iLet at non-default tmax settings. Improvements were observed in mean sensor glucose without increases in low sensor glucose at non-default tmax settings.
Trial Registration:
ClinicalTrials.gov, NCT03816761.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Oral Hypoglycemic Agents: Glinides
Dipeptidyl Peptidase 4 Inhibitors
Glucagon-like Receptor Agonists
GLP-1, when administered in high doses intravenously, triggers insulin secretion, inhibits glucagon release, slows gastric emptying, reduces food intake, and restores normal insulin secretion. However, its rapid inactivation by...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Insulin: Biosynthesis, Chemistry, and Preparation
Damage or functional impairment of β-cells inhibits insulin production, leading to diabetes. Diabetes treatment...