Validation of the inverse method for the determination of the access flow with thermodilution
M Luz Sánchez Tocino1, Antonio López González2, Silvia Villoria González1
1Fundación Renal Íñigo Álvarez de Toledo, Ciudad Rodrigo, Salamanca, España.
Introduction:
Thermodilution is a widely used method for measuring vascular access flow (QA). Among the possibilities of thermodilution, the reverse method (RM) can be beneficial in the execution time, without impact on the dialysis efficacy (Kt). However, it is not a sufficiently studied technique.
Method:
Transversal study of 117 arteriovenous fistulas. Two QA measurements were taken with the method described by the manufacturer (MR) and another with RM. RM is based on the obtention of an inverted recirculation registry at the beginning of the session and a single subsequent recirculation measurement with the lines in normal position. In the concordance analysis, the Bland-Altman method and Cohen's Kappa index were used.
Results:
Very good concordance between MR and RM was evidenced for QA below 700ml/min, but it worsens as flow increases. The median variability between the MR measurements (intra-method variability) was 3.4% (-17.13). This value did not differ from the median variability generated between MR and RM (inter-method variability), which was 2% (-14,12) (P=.287). The degree of agreement between the 2 to identify arteriovenous fistulas susceptible to intervention was very good (Kappa=0.834). The time spent using the RM was significantly shorter (P=.000) without evidence of variations in the Kt of the measurement sessions (P=.201).
Conclusions:
The thermodilution RM is valid to determine the flow of the vascular access, especially in QA lower than 700ml/min, with great time savings, simplification of the procedure and without modifying the dialysis efficiency. The variability between the measurement by MR and RM is similar to that of MR. The concordance between methods in identifying potentially pathological arteriovenous fistulas is very good.
Related Concept Videos
Pipe Flowrate Measurement: Problem Solving
Pipe Flowrate Measurement
The orifice meter is a simple,...
Joule-Thomson Effect
This experiment forces high-pressure gas through a throttle valve or a porous plug to a lower-pressure region. The gas expands as it passes through to...
Bernoulli's Equation: Problem Solving
The first step is to compute the cross-sectional areas of the pipe and the Venturi throat to analyze the pressure difference indicated by the pressure gauge. Next, the continuity...
Uniform Depth Channel Flow: Problem Solving
Constant Pressure Calorimetry