Virtual Unenhanced Images: Qualitative and Quantitative Comparison Between Different Dual-Energy CT Scanners in a
Simon Lennartz, Nisanard Pisuchpen, Anushri Parakh1
1From the Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, MA.
Materials And Methods:
Forty-four patients with clinical contrast-enhanced abdominal examinations on each of the 3 DECT scanner types and a phantom scanned with the same protocols were included in this retrospective study. Qualitative and quantitative assessment was performed on VUE images. Quantitative evaluation included measurement of attenuation and image noise for various tissues and the phantom. Virtual unenhanced image attenuation and noise were compared between scanner types, and intrapatient interscanner reproducibility of virtual unenhanced image attenuation was calculated as the percentage of measurement pairs with an interscanner difference ≤ 10 HU. Image quality, noise, sharpness, and iodine elimination were assessed qualitatively by 2 radiologists.
Results:
Significant interscanner differences in VUE attenuation and noise were found in all tissues. dlDECT and rsDECT showed significantly higher VUE attenuation than dsDECT in the aorta, portal vein, and kidneys (P < 0.05). Conversely, VUE attenuation in dsDECT was significantly higher than in dlDECT/rsDECT for subcutaneous and retroperitoneal fat (both P < 0.05). A total of 91.9% (385/419) of measurements were reproducible between rsDECT and dlDECT, 70.9% (297/419) between dsDECT and rsDECT, and 66.8% (280/419) between dsDECT and dlDECT. Virtual unenhanced image attenuation in the contrast media-filled phantom cavity was 12.7 ± 4.7 HU in dlDECT, -5.3 ± 4.2 HU in rsDECT, and -4.0 ± 10.7 HU in dsDECT with significant differences between dlDECT and rsDECT/dsDECT, respectively (P < 0.05), between which attenuation was comparable in the unenhanced extraluminal phantom component (P = 0.11-0.62). Qualitatively, dsDECT yielded best iodine elimination, whereas sharpness, image noise, and overall image quality were rated higher in dlDECT and rsDECT.
Conclusions:
There are significant interscanner differences in the attenuation measurements and qualitative assessment of VUE images, which should be acknowledged when using these images in patients that are being scanned on different DECT scanner types during imaging follow-up.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Ultrasonography