TFNA Implant Failure with Helical Blade Cut-out
Sanjay Agarwala1, Mayank Vijayvargiya1, Tej Shah1
1Department of Orthopedics, PD Hinduja Hospital and Medical Research Centre, Mumbai. Maharashtra. India.
This study reports the first case of Trochanteric Femoral Nail-Advanced (TFNA) helical blade cut-out in an elderly osteoporotic patient with an inter-trochanteric femur fracture. The failure was linked to a high tip-apex distance and poor bone quality, highlighting the need for careful fracture management.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
Background:
- The Trochanteric Femoral Nail-Advanced (TFNA) is a newer implant designed to improve upon the Proximal Femoral Nail Anti-rotation (PFNA).
- Despite design advancements, complications can still occur if fracture management principles are not strictly followed.
- This report details a rare complication associated with TFNA use.