Safety and Efficacy of Leadless Pacemakers: A Systematic Review and Meta-Analysis
Linh Ngo1,2,3, Daniel Nour2, Russell A Denman2
1School of Clinical Medicine The University of Queensland Brisbane QLD Australia.
Journal of the American Heart Association
|June 25, 2021
Summary
This systematic review found that the Micra leadless pacemaker has a low complication rate and good electrical performance up to one year. Further randomized trials are needed to confirm these findings for widespread adoption.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Leadless pacemakers represent a novel advancement in cardiac rhythm management.
- Evidence regarding the safety and efficacy of leadless pacemakers, particularly those implanted in the right ventricle, requires further elucidation.
Purpose of the Study:
- To conduct a systematic review and meta-analysis evaluating the safety and efficacy of leadless pacemakers.
- To specifically examine complication rates and pacing capture thresholds for Nanostim and Micra devices.
Main Methods:
- A comprehensive search of PubMed and Embase databases was performed for studies published before June 6, 2020.
- Included studies were observational, focusing on Nanostim and Micra leadless pacemakers.
- Primary safety outcome was major complications; primary efficacy endpoint was acceptable pacing capture threshold (≤2 V). Pooled estimates were calculated using the Freedman-Tukey double arcsine transformation.
Main Results:
- The Micra leadless pacemaker demonstrated a pooled complication incidence of 0.46% at 90 days and 1.77% at 1 year.
- Compared to transvenous pacemakers, Micra was associated with 51% lower odds of complications (OR, 0.49; 95% CI, 0.34-0.70).
- At 1 year, 98.96% of patients with Micra achieved good pacing capture thresholds. Nanostim showed higher complication rates and variable capture thresholds.
Conclusions:
- The Micra leadless pacemaker exhibits a favorable safety profile with low complication rates and effective electrical performance up to one year post-implantation.
- While promising, further data from randomized controlled trials are necessary to support the broad clinical adoption of leadless pacemakers.