Related Experiment Video
Updated: Nov 1, 2025

"Liver-on-a-Chip" Cultures of Primary Hepatocytes and Kupffer Cells for Hepatitis B Virus Infection
Published on: February 19, 2019
"Scratch it out": carbon copy based paper devices for microbial assays and liver disease diagnosis
Amrutha Hasandka1, Anusha Prabhu, Akshata Prabhu
1Department of Biotechnology, Manipal Institute of Technology, Manipal Academy of Higher Education, Manipal, Karnataka 576104, India. naresh.mani@manipal.edu maninaresh@gmail.com.
Abstract:
We present a facile paper-based microfluidic device fabrication technique leveraging off-the-shelf carbon paper for the deposition of hydrophobic barriers using a novel "stencil scratching" method. This exceedingly frugal approach (0.05$) requires practically no technical training to employ. Hydrophobic barriers fabricated using this approach offer a width of 3 mm and a hydrophilic channel width of 849 μm, with an ability to confine major aqueous solvents without leakage. The utility of the device is demonstrated by porting a cell viability assay showing a limit-of-detection (LOD) of 0.6 × 108 CFU mL-1 and bilirubin assay with human serum showing a detection range of 1.76-6.9 mg dL-1 and a limit-of-detection (LOD) of 1.76 mg dL-1. The intuitiveness and economic viability of the fabrication method afford it great potential in the field of point-of-care diagnostics geared towards providing testing infrastructure in resource-scarce regions globally.