Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
A Single-opening&closing Valve Tester for Direct Measurement of Closing Volume of the Heart Valve
Hao Wang1, Zhan Cui1, Zhongxi Zhou1
1Research Center of Fluid Machinery Engineering & Technology, Jiangsu University, Zhenjiang, 212013, Jiangsu Province, People's Republic of China.
Purpose:
The objective of this study was to develop a novel single opening&closing pulsatile flow in-vitro valve tester for direct measurement of closing volume of the heart valve.
Methods:
A single opening&closing valve tester was composed of a piston pump, valve mounting chamber, reservoir, measurement and control system. The piston pump was used to drive a valve to open and close with dictated flow which comprised three phases of accelerated, constant, and decelerated flow with six slopes. A high speed camera was used to record valve opening and closing images. Two pressure transducers across the tested valve were used to capture the ending time of valve closing which was verified by the high-speed photography. The closing time was measured and closing volume was calculated with a piston displacement volume during valve closing. A tilting disc valve and porcine mitral valve were tested.
Results:
There was a big difference in flowrate between the Transonic flowmeter and piston pump. The heart valve opened and closed under the dictated flow driven by the piston pump. The transvalvular pressure was minor during valve opening and then increased sharply during valve closing. The closing time varied approximately linearly with the slope of the decelerated flow and was comparable between the two methods by the transvalvular pressure and high-speed photography. The closing volumes did not change much with the slope of the decelerated flow and were 7.0 ± 1.0 and 14.0 ± 1.5 mL for the tilting disc valve and mitral valve, respectively.
Conclusion:
Pulsatile flow is challenging to the flowmeter. A novel single opening&closing pulsatile flow in-vitro valve tester for the heart valve has successfully been developed and can be used to simulate and evaluate the opening and closing hemodynamics of the heart valve. The tester can be used to measure valve closing volume and time accurately with a standardized testing protocol free from effect of other components such as the resistance, compliance units and auxiliary valve in the continuous pulsatile flow valve tester.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...