Related Experiment Video
Updated: Oct 31, 2025

Using a Cell-Tracer Injection to Investigate the Origin of Neointima-Forming Cells in a Rat Saccular Side Wall Model
Published on: March 16, 2022
Outcomes following aneurysmal coil embolization with intentionally shortened low-profile visible intraluminal support
Kenji Yatomi1, Yumiko Mitome-Mishima1,2, Takashi Fujii2,3
1Department of Neurosurgery, Juntendo University Faculty of Medicine, Japan.
Purpose:
Among all stents available for neuroendovascular therapy, the low-profile visible intraluminal support stent bears the highest metal coverage ratio. We deployed a low-profile visible intraluminal support stent with a delivery wire or/and microcatheter system push action to shorten the low-profile visible intraluminal support stent and thus achieve a flow diversion effect. We report our single-institution experience with the use of low-profile visible intraluminal support stents for intentionally shortened deployment (shortening group) and non-shortened deployment (non-shortening group) for unruptured intracranial aneurysms.
Methods:
We retrospectively reviewed the medical records of 130 patients with 131 intracranial aneurysms who were treated with low-profile visible intraluminal support stent-assisted coil embolization from February 2016-January 2019. All perioperative complications were noted. Every 6 months, we re-examined the patients with cerebral angiography or magnetic resonance angiography. The outcomes of aneurysm occlusion were evaluated by the modified Raymond-Roy occlusion classification. We used the finite element method and computational fluid dynamics to investigate the hemodynamics after shortened low-profile visible intraluminal support stent deployment.
Results:
Immediately after treatment, the modified Raymond-Roy occlusion classification was significantly better in the shortening group than in the non-shortening group (p<0.05). The latest angiographic outcomes showed the same tendency. Hemodynamic analysis by computational fluid dynamics suggested an adequate flow diversion effect with the use of our intentional shortening method.
Conclusions:
Stent-assisted coil embolization using this technique showed good results of a high complete occlusion rate and low complication rate. These findings suggest that shortened low-profile visible intraluminal support stent deployment yields a flow diversion effect and may lead to early intra-aneurysmal thrombus formation.