Related Experiment Video
Updated: Oct 30, 2025

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Biventricular Pacing Versus Right Ventricular Pacing in Patients Supported With LVAD
Ben B Chung1, Jonathan S Grinstein1, Teruhiko Imamura1
1Department of Medicine, University of Chicago, Chicago, Illinois, USA.
Objectives:
This study sought to evaluate the effects of right ventricular (RV) pacing versus biventricular (BiV) pacing on quality of life, functional status, and arrhythmias in LVAD patients.
Background:
Cardiac resynchronization therapy (CRT) and left ventricular assist devices (LVADs) independently improve outcomes in heart failure patients, but the effects of combining these therapies remains unknown. We present the first prospective randomized study evaluating the effects of RV versus BiV pacing on quality of life, functional status, and arrhythmias in LVAD patients.
Methods:
In this prospective randomized crossover study, LVAD patients with prior CRT devices were alternated on RV and BiV pacing for planned 7-14-day periods. Ambulatory step count, 6-minute walk test distance, Kansas City Cardiomyopathy Questionnaire scores, arrhythmia burden, CRT lead function, and echocardiographic data were collected with each pacing mode.
Results:
Thirty patients were enrolled, with a median age of 65 years, 67% male, and mean duration of LVAD support of 309 days. Compared with BiV pacing, RV-only pacing resulted in 29% higher mean daily step count, 11% higher 6-minute walk test distance, and 7% improved KCCQ-12 score (all p < 0.03). LV end-diastolic volume was significantly lower with RV pacing (220 vs. 250 mL; p = 0.03). Fewer patients had ventricular tachyarrhythmia episodes during RV pacing (p = 0.03). RV lead impedance was lower with RV pacing (p = 0.047), but no significant differences were observed in impedance across other CRT leads.
Conclusions:
In the first prospective randomized study comparing variable pacing in LVAD patients, RV pacing was associated with significantly improved functional status, quality of life, fewer ventricular tachyarrhythmias, and stable lead impedance compared with BiV pacing. This study supports turning off LV lead pacing in LVAD patients with CRT.
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiac Catheterization II: Right Heart Catheterization