Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
First-time evaluation of ascending compared to rectangular transthoracic defibrillation waveforms in modelled
Tobias Neumann1, Simon-Richard Finke1, Maja Henninger1
1University of Cologne, Faculty of Medicine and University Hospital Cologne, Department of Anaesthesiology and Intensive Care Medicine, Research Group "Clinical Research and Teaching", Kerpener Str. 62, 50937 Cologne, Germany.
Aim Of The Study:
Prognosis in out-of-hospital cardiac arrest (OHCA) depends on cardiopulmonary resuscitation (CPR) duration. Therefore, the optimal biphasic defibrillation waveform shows high conversion rates besides low energy. Matthew Fishler theoretically predicted it to be truncated ascending exponential. We realised a prototypic defibrillator and compared ascending with conventional rectangular waveforms in modelled OHCA and CPR.
Methods:
Approved by the authorities, 57 healthy swine (Landrace × Piétrain) were randomised to ASCDefib (n 26) or CONVDefib (n 26). Five swine served as sham control. We induced ventricular fibrillation (VF) electrically in anaesthetised swine randomised to ASCDefib or CONVDefib and discontinued mechanical ventilation. After 5 min of untreated cardiac arrest, we started CPR with mechanical chest compressions and ventilation. We performed transthoracic biphasic defibrillations after 2, 4, 6 and 8 min CPR targeting 4 J/kg in either group. Depending on the randomised group, the defibrillation protocol was either three ascending followed by one rectangular waveform (ASCDefib) or three rectangular followed by one ascending waveform (CONVDefib).
Results:
Under our model-specific conditions, VF was initially terminated by 13/80 ascending waveforms and 13/79 rectangular waveforms and persistent return of spontaneous circulation was achieved in 8/26 (ASCDefib) vs. 10/26 (CONVDefib) animals. Mean current rather than waveform design was predictive for defibrillation success in a generalised linear model.
Conclusion:
Contrary to theoretical assumptions, transthoracic biphasic defibrillation with ascending waveforms is not superior to rectangular waveforms in modelled OHCA. We advocate defibrillation dosage to be guided by current, that has proven its predictive value again.
Institutional Protocol Number:
84-02.04.2017.A176.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...