Related Experiment Video
Updated: Oct 29, 2025

In Vitro Thrombosis Test for Ventricular Assist Devices
Published on: March 21, 2025
A Power Tracking Algorithm for Early Detection of Centrifugal Flow Pump Thrombosis
Mark S Slaughter1, Thomas Schlöglhofer2, Jonathan D Rich3
1From the Department of Cardiovascular and Thoracic Surgery, University of Louisville, Louisville, Kentucky.
A new Power Tracking algorithm detects early signs of pump thrombosis (PT) in HeartWare HVAD System patients. This algorithm identifies deviations in pump power, enabling earlier medical intervention before critical events occur.
Area of Science:
- Cardiovascular Engineering
- Biomedical Data Science
- Medical Device Monitoring
Background:
- Logfiles from the HeartWare HVAD System offer crucial operational pump trend data for patient management.
- Pump thrombosis (PT) is a serious complication often indicated by increased logfile power preceding clinical symptoms.
Purpose of the Study:
- To develop and validate a Power Tracking algorithm for early detection of pump thrombosis (PT) using HVAD system logfiles.
- To assess the algorithm's sensitivity, timing of detection, and false alarm rate in clinical trial data.
Main Methods:
- A Power Tracking algorithm was developed to identify significant deviations in pump power indicative of PT.
- The algorithm was retrospectively applied to logfiles from the ENDURANCE, ENDURANCE Supplemental, and LATERAL clinical trials.
- Two cohorts were analyzed: suspected PT events (n=70) and control patients without adverse events (n=106).
Main Results:
- The Power Tracking algorithm detected PT cases on or before the recorded adverse event date with 85.7% sensitivity.
- Detection occurred an average of 3.9 days prior to clinical presentation.
- The algorithm generated approximately one false alarm per 6.85 patient-years of monitoring in non-event logfiles.
Conclusions:
- The Power Tracking algorithm demonstrates favorable performance for early PT detection in HVAD patients.
- Earlier detection facilitates timely medical management, potentially avoiding surgical intervention.
- This algorithm can enhance patient management by providing advance warning of potential pump thrombosis.