When Two Is Better than One : The Buddy-wire Technique in Flow-diversion Procedures
Tomas Dobrocky1,2, Hubert Lee3, Patrick Nicholson3
1Division of Neuroradiology, Joint Department of Medical Imaging, Toronto Western Hospital, Toronto, Ontario, Canada. tomas.dobrocky@insel.ch.
Background:
Delivery of most flow diverters (FD) requires larger, and thus stiffer microcatheters (0.021-0.027in.) which can pose challenges to intracranial navigation. The concomitant use of two microwires within one microcatheter, also known as the buddy-wire technique, may be helpful for navigation and support in challenging situations.
Methods:
We analyzed all flow diverter procedures in our prospectively collected database. We recorded all patient-related, anatomical and procedural information. We performed univariate statistics and technical descriptions.
Results:
In total, 208 consecutive patients treated with a FD at our institution between July 2014 and August 2020 were retrospectively analyzed. In 17 patients the buddy-wire technique was used (mean age 63 years, range 31-87 years: 16 female). Aneurysms were located at the petrous, cavernous, supraophthalmic internal carotid artery, and a proximal M2 branch in 2, 7, 7 and 1 patient(s), respectively. In all cases a 0.027in. microcatheter was used for device deployment. In 14 patients with a wide-necked aneurysm the buddy-wire provided additional support to advance the microcatheter and mitigated the ledge between the aneurysm neck and the parent artery or a side branch. In two giant cavernous aneurysms treated with telescoping FDs, the buddy-wire was used to re-enter the proximal end of the foreshortened FD.
Conclusion:
The buddy-wire is a useful technique in FD procedures to prevent herniation of the microcatheter into the aneurysm sack, in wide-necked aneurysms to mitigate the ledge effect between the aneurysm neck and the parent artery where the microcatheter tip may get stuck, or to enable re-entry into a foreshortened FD.