X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
Published on: August 20, 2019
Harmonics as an alternative method for measuring I0 during x-ray absorption spectroscopy experiments at laboratory
René Bes1, Saara Takala1, Simo Huotari1
1Department of Physics, University of Helsinki, P.O. Box 64, FI-00014 Helsinki, Finland.
This study introduces a new method for simultaneous x-ray absorption spectroscopy (XAS) measurements, simplifying laboratory-based experiments. This technique enhances the efficiency of XAS analysis using laboratory x-ray sources.
Area of Science:
- Materials Science
- Analytical Chemistry
- Spectroscopy
Background:
- Laboratory-scale spectrometers for X-ray Absorption Spectroscopy (XAS) are increasingly common.
- These instruments facilitate routine XAS studies, overcoming synchrotron access limitations.
- A key challenge in laboratory XAS is the need for separate measurements of incident (I₀) and transmitted (I₁) X-ray intensities due to lower source brightness.
Purpose of the Study:
- To present and validate an alternative method for simultaneous measurement of I₀ and I₁ in laboratory XAS.
- To demonstrate the reliability and robustness of this novel approach.
Main Methods:
- Utilized natural harmonics generated by monochromator crystals.
- Developed a simultaneous measurement technique for incident and transmitted X-ray intensities.
- Performed experimental validation at the Cobalt (Co) K-edge and Neodymium (Nd) L₃-edge.
Main Results:
- Successfully demonstrated simultaneous measurement of I₀ and I₁ using monochromator harmonics.
- Validated the approach with Co metal foil at the Co K-edge.
- Confirmed reliability using Nd₂O₃ at the Nd L₃-edge.
Conclusions:
- The proposed method offers a reliable and robust way to perform simultaneous intensity measurements in laboratory XAS.
- This advancement simplifies experimental procedures and enhances the utility of compact XAS systems.
- The technique leverages inherent properties of monochromator crystals for improved efficiency.
Related Concept Videos
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
UV–Vis Spectrometers
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...