Harmonics as an alternative method for measuring I0 during x-ray absorption spectroscopy experiments at laboratory scale
Related Experiment Video
Updated: Oct 29, 2025

X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
Published on: August 20, 2019
René Bes1, Saara Takala1, Simo Huotari1
1Department of Physics, University of Helsinki, P.O. Box 64, FI-00014 Helsinki, Finland.
Abstract:
In the recent years, the advent of an efficient and compact laboratory-scale spectrometer for x-ray absorption spectroscopy experiments has been extensively reported in the literature. Such modern instruments offer the advantage to routinely use x-ray absorption spectroscopy on systematic studies, which is usually unconceivable at synchrotron radiation source facilities due to often limited time access. However, one limiting factor is the fact that due to laboratory x-ray source brightness compared to a synchrotron, two separate measures of the incoming and transmitted x-ray intensities, i.e., the so-called I0 and I1, respectively, are usually required. Herein, we introduce and discuss an alternative approach for measuring I0 and I1 simultaneously. Based on the usage of harmonics arising naturally from the use of monochromator crystals, the reliability and robustness of our proposed approach is demonstrated through experiments at the Co K-edge measured using Co metal foil and at the Nd L3-edge measured in Nd2O3.
Related Concept Videos
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
UV–Vis Spectrometers
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...