Comparison of Lag Versus Nonlag Screw Fixation for Long Oblique Proximal Phalanx Fractures: A Biomechanical Study
Megan R Miles1, Tyler Green1, Brent G Parks1
1Curtis National Hand Center, MedStar Union Memorial Hospital, Baltimore, MD.
Purpose:
To compare lag versus nonlag screw fixation for long oblique proximal phalanx (P1) fractures in a cadaveric model of finger motion via the flexor and extensor tendons.
Methods:
We simulated long oblique P1 fractures with a 45° oblique cut in the index, middle, and ring fingers of 4 matched pairs of cadaveric hands for a total of 24 simulated fractures. Fractures were stabilized using 1 of 3 techniques: two 1.5-mm fully threaded bicortical screws using a lag technique, two 1.5-mm fully threaded bicortical nonlag screws, or 2 crossed 1.14-mm K-wires as a separate control. The fixation method was randomized for each of the 3 fractures per matched-pair hand, with each fixation being used in each hand and 8 total P1 fractures per fixation group. Hands were mounted to a custom frame where a computer-controlled, motor-driven, linear actuator powered movement of the flexor and extensor tendons. All fingers underwent 2,000 full flexion and extension cycles. Maximum interfragmentary displacement was continuously measured using a differential variable reluctance transducer. Our primary outcome was the difference in the mean P1 fragment displacement between lag and nonlag screw fixation at 2,000 cycles.
Results:
The observed differences in mean displacement between lag and nonlag screw fixation were not statistically significant throughout all time points. A two one-sided test procedure for paired samples confirmed statistical equivalence in the fragment displacement between these fixation methods at all time points, including the primary end point of 2,000 cycles.
Conclusions:
Nonlag screws provided equivalent biomechanical stability to lag screws for simulated long oblique P1 fractures during cyclic testing in this cadaveric model.
Clinical Relevance:
Fixation of long oblique P1 fractures with nonlag screws has the potential to simplify treatment without sacrificing fracture stability during immediate postoperative range of motion.