Multi-institutional TSA-amplified Multiplexed Immunofluorescence Reproducibility Evaluation (MITRE) Study
Janis M Taube1, Kristin Roman2, Elizabeth L Engle3
1Department of Dermatology, The Johns Hopkins Hospital, Baltimore, Maryland, USA jtaube1@jhmi.edu.
Multiplexed immunofluorescence (mIF) assays for PD-1/PD-L1 axis biomarkers show high reproducibility across multiple institutions. This standardized workflow enables reliable spatial analysis for precision immuno-oncology in clinical trials.
Area of Science:
- Oncology
- Immunology
- Biomarker Discovery
- Pathology
Background:
- Precision immuno-oncology relies on predictive biomarkers from spatial cell arrangements and coexpression patterns.
- Multiplexed immunofluorescence (mIF) is a key technology for assessing these biomarkers.
- Standardization of mIF workflows is crucial for multisite clinical trials and laboratory processes.
Purpose of the Study:
- To optimize and validate an automated six-plex mIF assay for the PD-1/PD-L1 axis.
- To assess the intersite and intrasite reproducibility of mIF staining and analysis.
- To evaluate the measurement of immune cell densities, PD-L1 expression, and PD-1/PD-L1 proximity.
Main Methods:
- A six-plex mIF panel (PD-L1, PD-1, CD8, CD68, FOXP3, CK) was optimized for quantitative equivalence to IHC.
- Six institutions stained serial tissue sections (tonsil, breast, NSCLC) using standardized automated stainers and imaging platforms.
- A locked-down image analysis algorithm was used to assess cell densities and coexpression patterns.
Main Results:
- The optimized mIF assay achieved 90% equivalence to IHC across six markers, with enhanced amplification for low-level expression.
- High intersite and intrasite concordance for cell density assessments were observed (R² values ranging from 0.72 to 0.88).
- Strong concordance was achieved for %PD-L1+ immune cells (R²=0.88) and PD-1/PD-L1 proximity (R²=0.82).
Conclusions:
- The optimized mIF assay provides sensitive and reproducible characterization of the PD-1/PD-L1 axis across multiple sites.
- High concordance across sites supports the use of mIF for assessing immune cell subsets, coexpression, and proximity with single-cell resolution.
- This standardized workflow is suitable for multisite clinical trials in precision immuno-oncology.