Related Experiment Video
Updated: Oct 28, 2025

Bone Marrow Transplantation Procedures in Mice to Study Clonal Hematopoiesis
Published on: May 26, 2021
Development of custom lead shield and strainer for targeted irradiation for mice in the gamma cell chamber
Nurhaslina Hasan1,2, Nur Fatihah Ronny Sham1, Muhammad Khalis Abdul Karim3
1Faculty of Medicine, Universiti Teknologi MARA, 47200, Sungai Buloh, Selangor, Malaysia.
Abstract:
We presented a development of a custom lead shield and mouse strainer for targeted irradiation from the gamma-cell chamber. This study was divided into two parts i.e., to (i) fabricate the shield and strainer from a lead (Pb) and (ii) optimize the irradiation to the mice-bearing tumour model with 2 and 8 Gy absorbed doses. The lead shielding was fabricated into a cuboid shape with a canal on the top and a hole on the vertical side for the beam path. Respective deliveries doses of 28 and 75 Gy from gamma-cell were used to achieve 2 and 8 Gy absorbed doses at the tumour sites.