Phase-contrast acceleration mapping with synchronized encoding
Simon Schmidt1,2,3, Martin Bruschewski4, Sebastian Flassbeck1,5,6
1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
Purpose:
To develop a phase-contrast (PC) -based method for direct and unbiased quantification of the acceleration vector field by synchronization of the spatial and acceleration encoding time points. The proposed method explicitly aims at in-vitro applications, requiring high measurement accuracy, as well as the validation of clinically relevant acceleration-encoded sequences.
Methods:
A velocity-encoded sequence with synchronized encoding (SYNC SPI) was modified to allow direct acceleration mapping by replacing the bipolar encoding gradients with tripolar gradient waveforms. The proposed method was validated in two in-vitro flow cases: a rotation and a stenosis phantom. The thereby obtained velocity and acceleration vector fields were quantitatively compared to those acquired with conventional PC methods, as well as to theoretical data.
Results:
The rotation phantom study revealed a systematic bias of the conventional PC acceleration mapping method that resulted in an average pixel-wise relative angle between the measured and theoretical vector field of (7.8 ± 3.2)°, which was reduced to (-0.4 ± 2.7)° for the proposed SYNC SPI method. Furthermore, flow features in the stenosis phantom were displaced by up to 10 mm in the conventional PC data compared with the acceleration-encoded SYNC SPI data.
Conclusions:
This work successfully demonstrates a highly accurate method for direct acceleration mapping. It thus complements the existing velocity-encoded SYNC SPI method to enable the direct and unbiased quantification of both the velocity and acceleration vector field for in vitro studies. Hence, this method can be used for the validation of conventional acceleration-encoded PC methods applicable in-vivo.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Relative Motion Analysis - Acceleration
Acceleration Vectors
Relative Motion Analysis using Rotating Axes - Acceleration
Time differentiation is...
Phase Changes
A substance melts or freezes at a temperature called its melting point and boils or condenses at its boiling point. These temperatures depend on pressure. High pressure favors the denser form of the substance, so typically, high pressure...
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...