Ventricular activation patterns during intrinsic conduction and right ventricular pacing in cardiac resynchronization
David O'Donnell1, Harish Manyam2, Carlo Pappone3
1Cardiology, GenesisCare, Wellington Parade, Melbourne, Australia.
Background:
Cardiac resynchronization therapy (CRT) involves stimulation of both right ventricle (RV) and left ventricle (LV). LV pacing from the sites of delayed electrical activation improves CRT response. The RV-LV conduction is typically measured in intrinsic rhythm. The differences in RV-LV conduction patterns and timing between intrinsic rhythm and during paced RV activation, these differences are not fully understood.
Methods:
Enrolled patients were implanted with a de novo CRT device and quadripolar LV lead, with lead implant locations at the implanting physician's discretion. QRS duration and conduction delay between the RV lead and each of the four LV electrodes (D1, M2, M3, and P4) were measured during intrinsic conduction and RV pacing.
Results:
Conduction measurements were collected from 275 patients across 14 international centers (68 ± 13 years of age, 73% male, 45% ischemic, 158 ± 22 ms QRS duration). Mean RV-LV conduction time was shorter during intrinsic conduction versus RV pacing by 59.6 ms (106.5 ± 36.5 versus 166.1 ± 32.1 ms, p < 0.001). The intra-LV activation delay between the latest and earliest activating LV electrode was also shorter during intrinsic conduction versus RV pacing by 6.6 ms (20.6 ± 13.1 vs. 27.2 ± 21.2 ms, p < 0.001). Intrinsic conduction and RV pacing resulted in a different activation order in 72.7% of patients, and the same LV activation order in 27.3%.
Conclusions:
Differences in RV-LV conduction time, intra-LV conduction time, and activation pattern were observed between intrinsic conduction and RV pacing. These findings highlight the importance of evaluating intrinsic versus paced ventricular activation to guide LV pacing site selection in CRT patients.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Electrophysiology of Normal Cardiac Rhythm
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias