Adverse events associated with femtosecond laser-assisted cataract surgery reported to the MAUDE database

Lisa R Koenig1, Andrew T Duong, Melissa Yuan

  • 1From the Department of Ophthalmology, Weill Cornell Medicine, New York, New York.

Abstract

Insights

Real-world data reveal common and rare complications of femtosecond laser-assisted cataract surgery (FLACS). Capsule tears and suction loss were most frequent, emphasizing the need for surgeon awareness and patient counseling regarding FLACS risks.

Area of Science:

  • Ophthalmology
  • Surgical Technology
  • Medical Device Safety

Background:

  • Femtosecond laser-assisted cataract surgery (FLACS) offers precision but carries potential complications.
  • Understanding real-world adverse events is crucial for patient safety and surgical technique refinement.

Purpose of the Study:

  • To comprehensively analyze and report the spectrum of real-world complications associated with femtosecond laser-assisted cataract surgery (FLACS).

Main Methods:

  • Retrospective review of deidentified data from the U.S. Food and Drug Administration's Manufacturer and User Facility Device Experience (MAUDE) database.
  • Analysis of adverse events involving ophthalmic femtosecond lasers during cataract surgery from January 2010 to January 2020.
  • Categorization of identified complications and trend analysis of event reporting over time.

Main Results:

  • A total of 1927 reports met inclusion criteria, detailing 2704 complications between 2010 and 2020.
  • The most frequent complications included capsule tears (41.2%), suction loss (13.7%), and capsulotomy-related issues (11.8%).
  • Corneal and incisional complications (16.0%) and rare events like miosis, endophthalmitis, and toxic anterior segment syndrome were also documented, with a peak in reported events in 2017.

Conclusions:

  • Real-world data from the MAUDE database highlight significant complications in FLACS, particularly capsule tears and suction loss.
  • Surgeons must be aware of the full range of potential FLACS complications to optimize patient preparation and informed consent.
  • Continuous monitoring of adverse events is essential for improving the safety and efficacy of femtosecond laser technology in cataract surgery.