Comparison of the expression of the volumetric alteration artifact in cylindrical and triangular fields of view in
Deivi Cascante-Sequeira1, Fernanda Coelho-Silva2, Lucas P Lopes Rosado2
1Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Av. Limeira, 901, PO Box 52, Piracicaba, SP, 13414-903, Brazil. dantikes@gmail.com.
Objective:
To compare the expression of the volumetric alteration (VA) artifact between cylindrical and convex triangular fields of view (FOVs) using high-density materials in different positions in two cone-beam computed tomography (CBCT) devices.
Materials And Methods:
Cylinders of five high-density materials (amalgam, chromium-cobalt, gutta-percha, titanium, and zirconium) with known physical volume were individually submitted to CBCT acquisition in four positions inside a polymethylmethacrylate phantom using two different FOV shapes (convex triangular and cylindrical) on the Veraviewepocs® R100 (R100) and Veraview® X800 (X800) devices. Two oral radiologists obtained the tomographic volumes by segmenting each cylinder. The difference between the tomographic and physical volumes corresponded to the VA. These values were analyzed by intraclass correlation coefficient and analysis of variance for repeated measures with Tukey post hoc test (α = 5%).
Results:
The FOV influenced the VA only in the X800 device (p = 0.014): the VA in the triangular FOV was greater than in the cylindrical FOV. The VA in the triangular FOV of the X800 device was greater than the R100 device (p < 0.0001). The material influenced the expression of the VA only in the R100 device (p < 0.0001); gutta-percha presented the highest VA, being underestimated, and differing from the other materials (p < 0.01).
Conclusion:
The triangular FOV increased the VA of high-density materials in the X800 device.
Clinical Relevance:
It is important to know if there is an influence on the volumetric alteration artifact of dental materials due to the different image formation geometry in the convex triangular FOV.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography