Novel label-free fluorescence aptasensor for chloramphenicol detection based on a DNA four-arm junction-assisted
Shuang Chen1, Yujie Liu1, Fei Zhai1
1Key Laboratory of Animal Resistance Biology of Shandong Province, Institute of Biomedical Sciences, Key Laboratory of Food Nutrition and Safety of Shandong Normal University, College of Life Science, Shandong Normal University, Jinan 250014, PR China.
Abstract:
A novel label-free fluorescence aptasensor was established for chloramphenicol (CAP) detection by DNA four-arm junction-assisted target recycling and SYBR Green I dye-aided fluorescence-signal amplification. The CAP aptamer was hybridized to its complementary strand (primer) to form a double-stranded primer/aptamer complex. In the presence of CAP, aptamers can specifically bind with CAP to dissociate primers, which can trigger the self-assembly of four hairpins to continuously generate DNA four-arm junctions. After digesting the excess hairpins using T7 exonuclease, SYBR Green I was inserted into the base pair-rich DNA four-arm junctions, which led to a significant increase in fluorescence intensity. Under optimal conditions, the developed aptasensor can detect CAP in a linear range of 1.0 pg mL-1 to 10 ng mL-1 with a detection limit of 0.72 pg mL-1. The recovery rates in milk and honey ranged from 90.3% to 106.6%. Thus, the method shows substantial potential for CAP detection in food products.