Related Experiment Video
Updated: Oct 26, 2025

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Amelioration of user experiences and glycaemic outcomes with an Advanced Hybrid Closed Loop System in a real-world
Pilar Isabel Beato-Víbora1, Fabiola Gallego-Gamero1, Ana Ambrojo-López1
1Endocrinology and Nutrition Department, Diabetes Technology Unit, Badajoz University Hospital, Badajoz, Spain.
Aims:
Automation in diabetes technology is rapidly evolving. The aim was to evaluate the real-world glycemic outcomes and user acceptance after 3 months of using the Medtronic 780G Advanced Hybrid Closed-Loop (AHCL) system.
Methods:
A prospective analysis was performed. A glucose target of 100 mg/dl and an active insulin time of 2 h were set. Capillary HbA1c, 2-week of pump and sensor data and several satisfaction questionnaire scores were compared at baseline and after 3 months of using the AHCL system.
Results:
52 subjects were selected (age: 43 ± 12 years, sex: 73% female, diabetes duration: 27 ± 11 years, higher education: 31%). Time in range (TIR) 70-180 mg/dl increased from 67.3 ± 13.6% to 80.1 ± 7.5% and time >180 mg/dl and >250 mg/dl were reduced (16.8 ± 8.4 vs 29.4 ± 15.1%, 2.7 ± 3.0% vs 6.9 ± 7.8%, respectively) (all p < 0.001), while time in hypoglycaemia remained below recommended targets. Time in Auto-Mode and sensor use were 94 ± 10% and 90 ± 11%, respectively. Auto-correction boluses represented 29 ± 12% of bolus insulin. Fear of hypoglycaemia, diabetes quality of life, sleep quality and satisfaction with the monitoring system improved after 3 months.
Conclusion:
The real-world use of the AHCL system Medtronic 780G provides an 80.1% TIR 70-180 mg/dl with minimal hypoglycaemia and an increased level of patient satisfaction.
Related Concept Videos
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Diabetes Mellitus: Type 2 and Gestational
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Oral Hypoglycemic Agents: Biguanides and Glitazones
Oral Hypoglycemic Agents: Glinides