Related Experiment Video
Updated: Oct 26, 2025

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Leadless left ventricular stimulation with WiSE-CRT System - Initial experience and results from phase I of SOLVE-CRT
Toshimasa Okabe1, John D Hummel1, Alan J Bank2
1The Ohio State University Wexner Medical Center, Columbus, Ohio.
Background:
Left ventricular (LV) endocardial pacing is a promising method to deliver cardiac resynchronization therapy (CRT). WiSE-CRT is a wireless LV endocardial pacing system, and delivers ultrasonic energy to an LV electrode.
Objective:
The purpose of this study was to present short-term outcomes with the WiSE-CRT system in centers with no prior implanting experience.
Methods:
Data were prospectively collected from 19 centers where WiSE-CRT systems were implanted during the roll-in phase of the SOLVE-CRT trial. Patients were followed at 1, 3, and 6 months, including transthoracic echo (TTE) at 6 months.
Results:
The WiSE-CRT was successfully implanted in all 31 attempted cases, and 30 patients completed the 6-month follow-up. One patient underwent heart transplantation 1 month after implantation, and was excluded. Fourteen (46.7%) patients demonstrated ≥1 NYHA class improvement. TTE data were available in 29 patients. LV ejection fraction, LV end-systolic volume, and LV end-diastolic volume improved from 28.3% ± 6.7% to 33.5% ± 6.9% (P < .001), 134.9 ± 51.3 mL to 111.1 ± 40.3 mL (P = .0004), and 185.4 ± 58.8 mL to 164.9 ± 50.6 mL (P = .0017), respectively. There were 3 (9.7%) device-related type 1 complications: 1 insufficient LV pacing, 1 embolization of an unanchored LV electrode, and 1 skin infection.
Conclusions:
We demonstrated a high success rate of LV endocardial electrode placement in centers with no prior implanting experience. Favorable clinical responses in heart failure symptoms and significant LV reverse remodeling were noted.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Acute Coronary Syndrome III: Diagnostic Studies
Cardiac Catheterization II: Right Heart Catheterization