Related Experiment Video
Updated: Oct 26, 2025

Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
Sensitive determination of gentamicin in plasma using ion-exchange solid-phase extraction followed by UHPLC-MS/MS
Ana Laura Anibaletto Dos Santos1,2, Anne Caroline Cezimbra da Silva1,2, Lilian de Lima Feltraco Lizot1,2
1Analytical Toxicology Laboratory, Universidade Feevale, Novo Hamburgo, RS, Brazil.
Background:
Therapeutic drug monitoring (TDM) of gentamicin sulfate (GEN) is usually recommended, particularly in critical patients. Only a few reports had described the determination of GEN in plasma or plasma using LC-MS/MS.
Objective:
This study aimed to develop and validate a sensitive ultra-high-performance liquid chromatography coupled to tandem mass spectrometry (UHPLC-MS/MS) assay for the quantification of GEN in small volumes of human plasma.
Results:
The use of a very low concentration of the ion-pairing agent HFBA allowed significant retention of the very polar GEN forms in a reversed phase UHPLC column. The solid-phase extraction (SPE) procedure allowed clean extracts, with no interferences detected in blank samples, and high sensitivity. The assay was linear on the range of 0.2-40 mg L-1 of GEN complex. The combined GEN complex had inter-assay CV of 8.8-10.0%, intra-assay CV of 10.2-11.0%, and accuracy of 96.8-104.0%. The assay was applied to 17 clinical samples obtained from neonate patients. Measured concentrations were in the range of 0.15-3.57 mg L-1 for GEN C1, 0.12-3.55 mg L-1 for GEN C1a, 0.20-5.77 mg L-1 for GEN C2, and 0.47-12.88 mg L-1 for the GEN complex, all within the linear range of the assay.
Conclusion:
A sensitive assay for the quantification of gentamicin in plasma using anion-exchange SPE and UHPLC-MS/MS was validated. The assay can be used for TDM of gentamicin, particularly in centers with access to proper instrumentation and with a low demand for gentamicin measurements, where immunoassays are not cost-effective.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
06:18A Novel Method to Determine the Longitudinal Antibacterial Activity of Drug-Eluting Materials
Published on: March 3, 2023
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Ion-Exchange Chromatography