Related Experiment Video
Updated: Oct 26, 2025

Assessment of Global Ocular Structure Following Spaceflight Using a Micro-Computed Tomography Micro-CT Imaging Method
Published on: October 27, 2020
MEASUREMENT OF RADIATION DOSE TO THE EYE LENS IN NON-ENHANCED CT SCANS OF THE BRAIN
Reem Alwasiah1, Abdulkarim Jawhari2, Rafal Al Orri2
1Department of Radiology, King Fahad General Hospital, Jeddah, Saudi Arabia.
Abstract:
Radiation can have undesirable effects on the eye, including a gradual loss of vision. Unnecessary radiation can reach the eye lens during non-enhanced computed tomography (CT) scans of the brain. The International Commission on Radiological Protection states the threshold for acute and chronic eye lens exposure is 500 mGy and the equivalent dose limit for the eye lens for public exposure is 15 mSv per year. Therefore, we measured the direct radiation dose to the eye lens during head CT scans using NanoDots in 216 adults. The mean absorbed dose to both eyes was 33.62 mGy (standard deviation ±12.442). The averages for the other variables were: tube current-time product: 260 mAs; dose-length product: 708 mGy cm and CT dose index: 35.5 mGy. Our findings encourage further study of radiation exposure and modifications in CT imaging protocols.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT