Related Experiment Video
Updated: Oct 25, 2025

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Balloon Seal Separation Leading to SAPIEN 3 Transcatheter Heart Valve Deployment Failure: Complications and
James Cockburn1, Sandeep Arunothayaraj1, Alex Gannaway1
1Sussex Cardiac Centre, University Hospitals Sussex NHS Trust, Brighton BN2 5DE, United Kingdom.
Abstract:
We present two cases of failure of balloon inflation secondary to balloon separation from the delivery catheter when implanting the SAPIEN 3 transcatheter heart valve (Edwards Lifesciences, Irvine, CA, USA). Although very uncommon, this is a potentially disastrous complication of transcatheter intervention. Case 1 highlights the complexity of the problem when it occurs and subsequent complications. Case 2 highlights how to manage this issue successfully.
Related Concept Videos
Mitral Stenosis III: Medical Management
Aneurysm III: Interprofessional Care
Aortic Regurgitation III: Medical Management
Endocarditis III: Medical Management
Mitral Regurgitation III: Medical Management
Mitral Stenosis IV: Nursing Management