Related Experiment Video
Updated: Oct 25, 2025

Thermal Ablation for the Treatment of Abdominal Tumors
Published on: March 7, 2011
Percutaneous lung and liver CT-guided ablation on swine model using microwave ablation to determine ablation size for
Paul Habert1,2,3, Mathieu Di Bisceglie1,2,3, Jean-François Hak1,2,3
1Department of Interventional Imaging, Assistance Publique Hopitaux de Marseille, Marseille, France.
Purpose:
Microwave ablation (MWA) provides an effective treatment of lung and liver tumors but suffers from a lack of reproducibility of ablation size among currently available technologies. In-vitro evaluations are far removed from clinical practices because of uninfused tissue. This study is in-vivo preclinical testing of a new MWA system on swine lungs and liver.
Materials And Methods:
All ablations were performed under CT guidance and multiple algorithms were tested with a power of 50, 75, and 100 W for durations of 3, 5, 8, 10, and 15 min. A 3 D-evaluation of the ablation zone was carried out using enhanced-CT. The sphericity index, coefficients of variation, and energy efficiency (which corresponds to the volume yield according to the power supplied) were calculated.
Results:
Fifty liver and 48 lung ablations were performed in 17 swine. The sphericity index varies from 0.50 to 0.80 for liver ablations and from 0.40 to 0.69 for lung ablations. The coefficient of variation was below 15% for 4/5 and 4/8 protocols for lung and liver ablations, respectively. The energy efficiency seems to decrease with the duration of the ablation from 0.60 × 10-3 cm3/J (75 W, 3 min) to 0.26 × 10-3 cm3/J (100 W, 15 min) in the liver and from 0.57 × 10-3 cm3/J (50 W, 10 min) to 0.42 × 10-3 cm3/J (100 W, 12 min) in the lungs.
Conclusion:
A shorter treatment time provides the best energy efficiency, and the best reproducibility is obtained for a 10 min treatment duration. The system tested provides an interesting reproducibility in both lung and liver measurements. Our results may help interventional radiologists in the optimal selection of treatment parameters.