Body-loop related MRI radiofrequency-induced heating hazards: Observations, characterizations, and recommendations
Xiaolin Yang1, Jianfeng Zheng1, Yu Wang1
1Department of Electrical and Computer Engineering, University of Houston, Houston, Texas, USA.
Purpose:
To assess RF-induced heating hazards in 1.5T MR systems caused by body-loop postures.
Methods:
Twelve advanced high-resolution anatomically correct human body models with different body-loop postures are created based on poseable human adult male models. Numerical simulations are performed to assess the radiofrequency (RF)-induced heating of these 12 models at 11 landmarks. A customized phantom is developed to validate the numerical simulations and quantitatively analyze factors affecting the RF-induced heating, eg, the contact area, the loop size, and the loading position. The RF-induced heating inside three differently posed phantoms is measured.
Results:
The RF-induced heating from the body-loop postures can be up to 11 times higher than that from the original posture. The RF-induced heating increases with increasing body-loop size and decreasing contact area. The magnetic flux increases when the body-loop center and the RF coil isocenter are close to each other, leading to increased RF-induced heating. An air gap created in the body loop or generating a polarized magnetic field parallel to the body loop can reduce the heating by a factor of three at least. Experimental measurements are provided, validating the correctness of the numerical results.
Conclusion:
Safe patient posture during MR examinations is recommended with the use of insulation materials to prevent loop formation and consequently avoiding high RF-induced heating. If body loops cannot be avoided, the body loop should be placed outside the RF transmitting coil. In addition, linear polarization with magnetic fields parallel to the body loop can be used to circumvent high RF-induced heating.
More Related Videos
09:01Magnetic-, Acoustic-, and Optical-Triple-Responsive Microbubbles for Magnetic Hyperthermia and Pothotothermal Combination Cancer Therapy
Published on: May 22, 2020
08:52Near-Infrared Temperature Measurement Technique for Water Surrounding an Induction-heated Small Magnetic Sphere
Published on: April 30, 2018
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...