Evaluation of Acuros® XB accuracy for static small fields dose calculations based on the IAEA/AAPM TRS-483
José M Lárraga-Gutiérrez1, Olivia A García-Garduño1, José A Herrera-González2
1Lab. de Física Médica, Instituto Nacional de Neurología y Neurocirugía, Insurgentes sur 3877, La Fama, Tlalpan 14269, CMDX, Mexico.
Purpose:
Evaluate Acuros® XB dose calculation accuracy following TRS-483 recommendations in small static fields for flattened and un-flattened 6 MV X-ray beams.
Methods:
Field output factors were measured following TRS-483 recommendations using four radiation detectors. Two sets of field output factors were measured. One set was used to configure the beam model into Acuros® XB down to a jaw-defined field size of 1.0 cm × 1.0 cm. The second set was used to evaluate the differences between calculated and measured field output factors for MLC-fields down to a field size of 0.5 cm × 0.5 cm.
Results:
Acuros® XB showed an accuracy within 1.5% down to an MLC-field of 1.0 cm × 1.0 cm, for a focal spot size of 1.0 and 0.0 mm in the cross and in-plane directions. For an MLC-field of 0.5 cm × 0.5 cm, an agreement was found within 3% between calculated and measured field output factors. These results were addressed by optimizing the focal spot size to minimize the differences between calculated and measured dose profiles.
Conclusions:
By optimizing the focal spot size, Acuros® XB showed an acceptable agreement within 3% down to an MLC-field of 0.5 cm × 0.5 cm. The results of this work suggest that if static and modulated delivery of very small targets is planned, then a field output factor table down to a field size of 1.0 cm is required in the beam configuration model.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...