Fabrication of microcalcifications for insertion into phantoms used to evaluate x-ray breast imaging systems
Bahaa Ghammraoui1, Ahmed Zidan2, Alaadin Alayoubi2
1Division of Imaging, Diagnostics, and Software Reliability, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD 20993, United States of America.
Abstract:
Physical breast phantoms can be used to evaluate x-ray imaging systems such as mammography, digital breast tomosynthesis and dedicated breast computed tomography (bCT). These phantoms typically attempt to mimic x-ray attenuation properties of adipose and fibroglandular tissues within the breast. In order to use these phantoms for task-based objective assessment of image quality, relevant diagnostic features should be modeled within the phantom, such as mass lesions and/or microcalcifications. Evaluating imaging system performance in detecting microcalcifications is of particular interest due to its' clinical significance. Many previously-developed phantoms have used materials that model microcalcifications using unrealistic chemical composition, which do not accurately portray their desired x-ray attenuation and scatter properties. We report here on a new method for developing real microcalcification simulants that can be embedded in breast phantoms. This was achieved in several steps, including cross-linking hydroxyapatite and calcium oxalate powders with a binder called polyvinylpyrrolidone (PVP), and mechanical compression. The fabricated microcalcifications were evaluated by measuring their x-ray attenuation and scatter properties using x-ray spectroscopy and x-ray diffraction systems, respectively, and were demonstrated with x-ray mammography and bCT images. Results suggest that using these microcalcification models will make breast phantoms more realistic for use in evaluating task-based detection performance of the abovementioned breast imaging techniques, and bode well for extending their use to spectral imaging and x-ray coherent scatter computed tomography.