Related Experiment Video
Updated: Oct 23, 2025

Ultrasensitive Detection of Biomarkers by Using a Molecular Imprinting Based Capacitive Biosensor
Published on: February 16, 2018
Molecularly Imprinted Polymer-Based Quartz Crystal Microbalance Sensor for the Clinical Detection of Insulin
Duygu Çimen1, Nilay Bereli1, Fatma Kartal1
1Department of Chemistry, Hacettepe University, Ankara, Turkey.
Abstract:
In this study, we reported the design of a quartz crystal microbalance (QCM) sensors for selective insulin detection. In the first step, N-methacryloyl-(L) 3-histidine methyl ester (MAH) monomer was formed a complex with insulin. Then, 2-hydroxyethyl methacrylate and ethylene glycol dimethacrylate were mixed with MAH:insulin complex. Insulin-imprinted and non-imprinted QCM sensors were synthesized by ultraviolet polymerization for the insulin detection. Insulin-imprinted QCM sensors was characterized by the contact angle measurements, atomic force microscopy and ellipsometry. Limit of detection (LOD) was found as 0.00158 ng/mL for the insulin-imprinted QCM sensors. Selectivity of insulin-imprinted and non-imprinted QCM sensors was carried in the presence of glucagon and aprotinin. Insulin-imprinted QCM sensor for insulin detection was also examined in the artificial plasma.