Related Experiment Videos
[Evaluation of a colorimetric enzymatic method for the determination of sialic acid in serum]
R Grisler1, C Santeramo, C Giavardi
1Laboratorio di ricerche cliniche, anatomia ed istologia patologica, Istituti Clinici di Perfezionamento, Milano.
Abstract:
Evaluation of sialic acid in serum was done by means of an enzymatic colorimetric method using the Biochemia Sialic Acid kit. Precision and accuracy within-run and run-to-run were high and linearity was observed just to a 170 mg/dl concentration. Reference values evaluated in 2700 samples were: mean = 65.3 mg/dl; mean +/- 2 standard deviations = 52.0-78.6 mg/dl.