Related Experiment Video
Updated: Oct 23, 2025

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
[Investigation of Verification Methods of Field Matching Using Polymer Gel Dosimeter in Proton Therapy]
Masaya Ishida1, Hiraku Kawamura2, Masaru Sato1
1Proton Medical Research Center, University of Tsukuba.
Purpose:
In our proton beam therapy center, we use imaging plates (IP) for dose verification of field matching in irradiation (called patch-field technique). In this study, a polymer gel dosimeter, which can perform three- dimensional dose distribution measurement, was used as a new tool for the verification of the patch field irradiation method corresponding to the rectangular irradiation.
Method:
The results of measurements of the PAGAT gel dosimeter to irradiate to two rectangular fields called as patch irradiation fields, which were created using the treatment planning system, were evaluated and compared to the results of IP and plan in profiles near the boundary of two fields.
Result:
In the case of no gap between the two fields, the relative dose using the gel dosimeter was 10.1% higher compared to that measured with the IP in the midpoint of two fields. In case of overlap (called hot region), the result of gels was 6.3% higher than that of IP. In the case of space (called cold region), the result of gels was 14.9% higher than that of IP. The results of the difference between gel and plan in the midpoint of two fields were 14.2% (no gap), -5.0% (hot), and 10.5% (cold).
Conclusion:
We found that the gel dosimeter was a 3-D dosimetric tool and possibility method for dose verification of patch fields. In this study, the results were preliminary and included several error factors. In the future, it is necessary to develop a dosimeter with improved and more precise measurements.

