Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
SURFACE DOSE ESTIMATION BY A KAP METER FOR KILOVOLTAGE X-RAY BEAMS
1Department of Health Sciences, Faculty of Life Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976, Japan.
Abstract:
This study aims to estimate the entrance surface dose (ESD) of a water phantom for kilovoltage x-ray beams using an air kerma area product meter (KAP meter) equipped in an x-ray unit. The KAP meter was calibrated in terms of the ESD determined by a plane-parallel ionization chamber based on a 60Co absorbed dose-to-water calibration coefficient, ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$. The ESD measured using the KAP meter was verified by comparing it with that estimated by the air kerma calibration coefficient, NK, for x-ray beam qualities. The ratio of ESDs based on ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$ and NK was 1.003 on average and independent of the beam quality. The ESD by the KAP meter was an agreement within ±1.5% with that measured using the plane-parallel chamber for 10 × 10-30 × 30 cm2 fields with a source-surface distance of 75-150 cm. It was possible to estimate the ESD directly in a water phantom for x-ray beams without correction factors compared to the existing air kerma calibration, using a KAP meter calibrated based on ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:23Author Spotlight: A Machine-Vision Approach to Transmission Electron Microscopy Workflows, Results Analysis and Data Management
Published on: June 23, 2023
Related Concept Videos
Biological Effects of Radiation
One-Compartment Open Model: Wagner-Nelson and Loo Riegelman Method for ka Estimation
On...
Dose Size and Dosing Frequency: Determination Methods
X-ray Imaging
Absorption of Radiation