Optimization of pseudo-continuous arterial spin labeling at 7T with parallel transmission B1 shimming
Kai Wang1, Samantha J Ma1,2, Xingfeng Shao1
1Laboratory of FMRI Technology, USC Mark & Mary Stevens Neuroimaging and Informatics Institute, Keck School of Medicine, University of Southern California, Los Angeles, California, USA.
Purpose:
To optimize pseudo-continuous arterial spin labeling (pCASL) for 7 T, and to further improve the labeling efficiency with parallel RF transmission transmit B1 ( ) shimming.
Methods:
pCASL parameters were optimized based on field distributions at 7 T with simulation. To increase labeling efficiency, the amplitude at inflowing arteries was increased with parallel RF transmission shimming. The "indv-shim" with shimming weights calculated for each individual subject, and the "univ-shim" with universal weights calculated on a group of 12 subjects, were compared with circular polarized (CP) shim. The optimized pCASL sequences with three shimming modes (indv-shim, univ-shim, and CP-shim) were evaluated in 6 subjects who underwent two repeated scans 24 hours apart, along with a pulsed ASL sequence. Quantitative metrics including mean amplitude, perfusion, and intraclass correlation coefficient were calculated. The optimized 7T pCASL was compared with standard 3T pCASL on 5 subjects, using spatial SNR and temporal SNR.
Results:
The optimal pCASL parameter set (RF duration/gap = 300/250 us, ) achieved robust perfusion measurement in the presence of inhomogeneities. Both indv-shim and univ-shim significantly increased amplitude compared with CP-shim in simulation and in vivo experiment (P < .01). Compared with CP-shim, perfusion signal was increased by 9.5% with indv-shim (P < .05) and by 5.3% with univ-shim (P = .35). All three pCASL sequences achieved fair to good repeatability (intraclass correlation coefficient ≥ 0.5). Compared with 3T pCASL, the optimized 7T pCASL achieved 78.3% higher spatial SNR and 200% higher temporal SNR.
Conclusion:
The optimized pCASL achieved robust perfusion imaging at 7 T, while both indv-shim and univ-shim further increased labeling efficiency.