[Pre-analytical phase impact in the differences in natremia obtained by direct and indirect ion selective electrodes
Antoine Puravet1, Claire Dupuis2, Charlotte Oris1
1Service de biochimie et génétique moléculaire, CHU Clermont-Ferrand, France.
Abstract:
Natremia is an important biological parameter providing information on the hydration state of patient's intracellular sector. Its measurement can be carried out either by multiparametric laboratory analyser (indirect potentiometry) or delocalized biology analyser (direct potentiometry). The main problem is that for a same patient, these two analysers can give quite different results, hence inducing interpretation problems for clinician. Two one-week study periods comparing the variations in blood sodium levels produced by these automatic analysers were carried out in two intensive care units of Clermont-Ferrand University Hospital. During the second study period, a protocol for collecting blood samples was applied in order to improve the pre-analytical conditions. Between the two weeks of studies, the median of the differences in natremia was significantly reduced, going from 4 mmol/L to 2 mmol/L (p < 0.001), as was the proportion of patients with large differences in sodium levels (strictly higher than 3 mmol/L) going from 51% to 24.8% (p < 0.001). The patients still presenting large variations in sodium had a median of proteins significantly lower than patients with deviations less than or equal to 3 mmol/L: 58.1 g/L against 62.25 g/L respectively (p < 0.001) leading to pseudo-hypernatremia (indirect potentiometry). Despite a significant reduction in differences linked to the application of good preanalytical practices, some patients nonetheless presented a major difference in natremia due to the difference of technique (variations in the lipidoprotein phase of the plasma of intensive care patients) and to the measurement uncertainties.
More Related Videos
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
Electrogravimetric Analysis: Overview
To test the completeness of the...
Ion-Exchange Chromatography
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Overview