Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
Ventilator output splitting interface 'ACRA': Description and evaluation in lung simulators and in an experimental
Pablo E Otero1, Lisa Tarragona1, Andrea S Zaccagnini1
1Cátedra de Anestesiología y Algiología, Facultad de Ciencias Veterinarias, Buenos Aires, Argentina.
Abstract:
The current COVID-19 pandemic has led the world to an unprecedented global shortage of ventilators, and its sharing has been proposed as an alternative to meet the surge. This study outlines the performance of a preformed novel interface called 'ACRA', designed to split ventilator outflow into two breathing systems. The 'ACRA' interface was built using medical use approved components. It consists of four unidirectional valves, two adjustable flow-restrictor valves placed on the inspiratory limbs of each unit, and one adjustable PEEP valve placed on the expiratory limb of the unit that would require a greater PEEP. The interface was interposed between a ventilator and two lung units (phase I), two breathing simulators (phase II) and two live pigs with heterogeneous lung conditions (phase III). The interface and ventilator adjustments tested the ability to regulate individual pressures and the resulting tidal volumes. Data were analyzed using Friedman and Wilcoxon tests test (p < 0.05). Ventilator outflow splitting, independent pressure adjustments and individual tidal volume monitoring were feasible in all phases. In all experimental measurements, dual ventilation allowed for individual and tight adjustments of the pressure, and thus volume delivered to each paired lung unit without affecting the other unit's ventilation-all the modifications performed on the ventilator equally affected both paired lung units. Although only suggested during a dire crisis, this experiment supports dual ventilation as an alternative worth to be considered.
Related Concept Videos
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Mechanical Ventilation I: Indication and Settings
Acute Respiratory Failure-III