Related Experiment Video
Updated: Oct 22, 2025

Electrochemical Preparation of Poly3,4-Ethylenedioxythiophene Layers on Gold Microelectrodes for Uric Acid-Sensing Applications
Published on: July 28, 2021
Developing Activated Carbon Veil Electrode for Sensing Salivary Uric Acid
Maria A Bukharinova1, Natalia Yu Stozhko1, Elizaveta A Novakovskaya1
1Scientific and Innovation Center of Sensor Technologies, Department of Physics and Chemistry, Ural State University of Economics, 8 Marta St., 62, 620144 Yekaterinburg, Russia.
A new carbon veil-based electrode (CVE) was developed for sensitive uric acid (UA) detection in saliva. Electrochemical activation enhances surface properties, enabling accurate and cost-effective UA measurements.
Area of Science:
- Electrochemistry
- Analytical Chemistry
- Materials Science
Background:
- Uric acid (UA) is a significant component of saliva, contributing to its antioxidant activity (AOA).
- Accurate and rapid determination of UA in saliva is crucial for health monitoring.
- Existing methods for UA detection may lack sensitivity, scalability, or cost-effectiveness.
Purpose of the Study:
- To develop a novel carbon veil-based electrode (CVE) for sensitive and selective determination of uric acid (UA) in saliva.
- To investigate the effect of electrochemical activation on the electrode's surface properties and performance.
- To evaluate the sensor's analytical characteristics and potential for practical application in saliva analysis.
Main Methods:
- Fabrication of carbon veil-based electrodes (CVE) using lamination technology.
- Electrochemical activation of CVEs via potentiostatic polarization in H2SO4 solution.
- Characterization using energy dispersive X-ray spectroscopy (EDS), electrochemical impedance spectroscopy (EIS), chronoamperometry, and linear sweep voltammetry (LSV).
Main Results:
- Electrochemical activation (CVEact) significantly increased surface area (1.5x), oxygen/nitrogen groups, and reduced charge transfer resistance.
- CVEact lowered UA oxidation potential by over 0.4 V compared to non-activated CVE.
- The developed sensor exhibited a low detection limit (0.05 µM), wide linear range (0.09-700 µM), high recovery (96-105%), and low RSD (≤3.9%).
Conclusions:
- The electrochemically activated carbon veil-based electrode (CVEact) is a highly sensitive and effective sensor for UA determination in saliva.
- The simple, scalable, and cost-effective manufacturing process makes CVEact suitable for widespread application.
- The sensor demonstrates potential for quick and reliable assessment of UA levels and its contribution to saliva's antioxidant activity.
More Related Videos
10:02Quantitative SERS Detection of Uric Acid via Formation of Precise Plasmonic Nanojunctions within Aggregates of Gold Nanoparticles and Cucurbit[n]uril
Published on: October 3, 2020
08:19Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012