Related Experiment Video
Updated: Oct 22, 2025

Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Agreement Between StepWatch3 and ActiGraph wGT3X+ for Measuring Step-Based Metrics in People With Peripheral Artery
Abstract:
The authors investigated the agreement between StepWatch3™ (SW3) and ActiGraph™ wGT3X+ monitors for measuring step-based metrics in patients with peripheral artery disease and older adults. In 23 patients with peripheral artery disease and 38 older participants, the authors compared the metrics obtained during an outdoor (400-m track) walking session (step count) and a 7-day free-living period (step count and 60/30/5/1-min maximal or peak step accumulation) using the SW3 (ankle) and the wGT3X+ (hip) with the low-frequency extension filter enabled (wGT3X+/LFE) or not (wGT3X+/N). During outdoor walking session, agreement was high, particularly for wGT3X+/LFE: correlations ≥.98, median absolute percentage errors <1%, and significant equivalence using a ± 15% equivalence zone or narrower. In free living, no wGT3X+ method was equivalent to SW3 for step count. The wGT3X+/LFE was equivalent to SW3 regarding all step accumulation metrics using a ± 20% equivalence zone or narrower, with median absolute percentage errors <11%. The wGT3X+/LFE method is the best option for comparisons with SW3 in peripheral artery disease and older adults.
Related Concept Videos
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessment of blood pressure in brachial artery(two-step method)
Peripheral Artery Disease I: Introduction