Related Experiment Video
Updated: Jul 2, 2026

Assessment and Evaluation of the High Risk Neonate: The NICU Network Neurobehavioral Scale
Published on: August 25, 2014
Newborn Incubators Do Not Protect from High Noise Levels in the Neonatal Intensive Care Unit and Are Relevant Noise
Tanja Restin1,2, Mikael Gaspar2, Dirk Bassler1
1Department of Neonatology, Newborn Research Zurich, University Hospital Zurich, 8091 Zurich, Switzerland.
Background:
While meaningful sound exposure has been shown to be important for newborn development, an excess of noise can delay the proper development of the auditory cortex.
Aim:
The aim of this study was to assess the acoustic environment of a preterm baby in an incubator on a newborn intensive care unit (NICU).
Methods:
An empty but running incubator (Giraffe Omnibed, GE Healthcare) was used to evaluate the incubator frequency response with 60 measurements. In addition, a full day and night period outside and inside the incubator at the NICU of the University Hospital Zurich was acoustically analyzed.
Results:
The fan construction inside the incubator generates noise in the frequency range of 1.3-1.5 kHz with a weighted sound pressure level (SPL) of 40.5 dB(A). The construction of the incubator narrows the transmitted frequency spectrum of sound entering the incubator to lower frequencies, but it does not attenuate transient noises such as alarms or opening and closing of cabinet doors substantially. Alarms, as generated by the monitors, the incubator, and additional devices, still pass to the newborn.
Conclusions:
The incubator does protect only insufficiently from noise coming from the NICUThe transmitted frequency spectrum is changed, limiting the impact of NICU noise on the neonate, but also limiting the neonate's perception of voices. The incubator, in particular its fan, as well as alarms from patient monitors are major sources of noise. Further optimizations with regard to the sound exposure in the NICU, as well as studies on the role of the incubator as a source and modulator, are needed to meet the preterm infants' multi-sensory needs.