Extension and improvement of the methanol-d4 NMR thermometer calibration
Niels Karschin1, Stephan Krenek2, Dieter Heyer2
1Department of NMR Based Structural Biology, Max Planck Institute for Biophysical Chemistry, Göttingen, Germany.
Abstract:
NMR thermometers are a convenient way to determine the temperature inside the sample of an NMR spectrometer. They rely on signals with strongly temperature-dependent chemical shifts, often of OH groups; 99.8% perdeuterated methanol is an established example which is particularly well suited for modern, high-sensitivity spectrometers, but it is so far calibrated only in the range of 282 to 330 K. In this work, we extend this calibration to the entire liquid range of methanol, 175 to 338 K. Additionally, we use a temperature sensor calibrated traceably to the International Temperature Scale (ITS-90) and accounted for the magnetic field effect on the sensor, yielding a more accurate calibration curve with an uncertainty (2σ) varying between 25 and 190 mK.
More Related Videos
Related Concept Videos
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
¹H NMR of Conformationally Flexible Molecules: Variable-Temperature NMR
NMR Spectrometers: Resolution and Error Correction
¹H NMR of Labile Protons: Deuterium (²H) Substitution
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...