Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Object size determination at computed tomography
1Department of Diagnostic Radiology, Uppsala University, Sweden.
Abstract:
To investigate whether large differences in attenuation between an object and its background at CT examination alter the apparent size of the object on the image compared with the conditions with a small attenuation difference, phantom experiments were performed. At the same time the influence of large attenuation differences on the geometric resolution was examined. The results show that neither the size reproduction of the object nor the geometric resolution is altered by variations in attenuation differences. Especially with large differences in attenuation between the object and background, however, the window centre setting was of greatest importance for correct reproduction of the object.