Feasibility of portable capnometer for mechanically ventilated preterm infants in the delivery room

Masashi Hotta1, Katsuya Hirata2, Masatoshi Nozaki2

  • 1Department of Neonatal Medicine, Osaka Women's and Children's Hospital, 840 Murodo-cho, Izumi, Osaka, 594-1101, Japan. spyf88d9@yahoo.co.jp.

This study aimed to determine whether a specific portable capnometer (EMMA™) can facilitate the maintenance of an appropriate partial pressure of arterial carbon dioxide (PaCO2) in intubated preterm infants in the delivery room. This study included preterm infants with a gestational age of 26 + 0 to 31 + 6 weeks who required intubation in the delivery room. We prospectively identified 40 infants who underwent the EMMA™ monitoring intervention group and 43 infants who did not undergo monitoring (historical control group). PaCO2 was evaluated either at admission in the neonatal intensive care unit or at 2 h after birth. The proportion of infants with an appropriate PaCO2 (35-60 mmHg) was greater in the intervention group than in the control group (80% vs. 42%; p = 0.001). There were no significant differences in the rate of accidental extubation (5.0% vs. 7.0%, p = 1.00) or in the proportion of infants with an appropriate PaCO2 among infants whose birth weight was < 1000 g (54% vs. 40%, p = 0.49). However, among infants whose birth weight was ≥ 1000 g, the PaCO2 tended to be more appropriate in the intervention group than in the control group (93% vs. 44%; p < 0.001).Conclusion: The EMMA™ facilitated the maintenance of an appropriate PaCO2 for mechanically ventilated preterm infants, especially infants with birth weight ≥1000 g, in the delivery room. What is Known: • An inappropriate partial pressure of arterial carbon dioxide has been associated with intraventricular hemorrhage in preterm infants. • There is a need to appropriately control the partial pressure of arterial carbon dioxide in preterm infants. What is New: • This is the first report regarding the feasibility of a portable capnometer, the EMMA™, in the delivery room. • The EMMA™ may be considered a feasible monitoring device in the delivery room for intubated preterm infants, especially infants with birth weight ≥1000 g.

Related Concept Videos

Assessment of Ventilation I: Respiratory Rate01:20

Assessment of Ventilation I: Respiratory Rate

Assessment of Ventilation
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
1.4K
Temperature Measurement Sites01:14

Temperature Measurement Sites

A thermometer measures body temperature. The common sites for measuring body temperature are the oral cavity, axillary region, temporal artery, and skin surface, such as the forehead, abdomen, and axilla. True core body temperature is assessed in the rectum, tympanic membrane, pulmonary artery, esophagus, and urinary bladder.
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
2.4K
Assessment of Diffusion and Perfusion01:17

Assessment of Diffusion and Perfusion

Understanding and evaluating diffusion and perfusion is critical in assessing a patient's respiratory and circulatory health. These processes play key roles in maintaining the body's internal environment, ensuring that tissues receive adequate oxygen while waste products are efficiently removed.
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
1.2K
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen01:16

Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen

Oxygen therapy is a pivotal aspect of medical care, particularly for patients with respiratory ailments. Two prominent oxygen-delivering systems include the Venturi mask and the transtracheal oxygen catheter.
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
1.1K
Mechanical Ventilation III: Noninvasive Ventilation01:23

Mechanical Ventilation III: Noninvasive Ventilation

Noninvasive positive-pressure ventilation (NIPPV), continuous positive airway pressure (CPAP), and bilevel positive airway pressure (BiPAP) are essential methods in respiratory care. These ventilation techniques offer unique benefits for patients with various respiratory conditions, providing adequate support without requiring intubation. Let's explore how each method is crucial in improving patient outcomes and enhancing respiratory therapy.
Noninvasive Positive-Pressure Ventilation...
325
Mechanical Ventilation I: Indication and Settings01:29

Mechanical Ventilation I: Indication and Settings

Mechanical ventilation is a life-saving technique for managing acute respiratory failure and other respiratory complications. The process involves using a machine known as a ventilator to supply oxygen to the lungs and assist in removing carbon dioxide. It serves as a bridge to long-term mechanical ventilation or a temporary measure until ventilatory support is discontinued. The ventilator can maintain this function for a prolonged period, providing critical support for patients until they can...
1.4K