Related Experiment Video
Updated: Oct 20, 2025

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
A New Solution for Stenting Large Right Ventricular Outflow Tracts Before Transcatheter Pulmonary Valve Replacement
Raymond N Haddad1, Damien Bonnet2, Ibrahim A Abu Zahira1
1Centre de Référence Malformations Cardiaques Congénitales Complexes-M3C, Hôpital Universitaire Necker-Enfants Malades, AP-HP, Paris, France.
Background:
Prestenting right ventricular outflow tracts (RVOTs) before transcatheter pulmonary valve replacement (TPVR) is essential. Optimus-XXL (AndraTec GmbH, Koblenz, Germany) is a new extra-large, balloon-expandable cobalt-chrome stent with promising technologies.
Methods:
From June 2020 to November 2020, 15 patients with congenital heart disease, dysfunctional RVOTs and target TPVR diameter ≥ 23 mm received Optimus-XXL stents before proceeding to TPVR using the SAPIEN valve (Edwards Lifesciences, Irvine, CA). Standard safety and outcomes were prospectively assessed.
Results:
Patients' median age and weight were 25.8 years (range: 10.5-63.1 years) and 58 kg (range: 43.8-101 kg), respectively. Underlying diagnosis was tetralogy of Fallot (66.7%), and RVOTs were patched (80%). Fifteen bare-metal stents were implanted using femoral (n = 14) and jugular approaches (n = 1). One conduit rupture was immediately controlled with a covered Optimus-XXL. Median stent length was 43 mm (range: 33-57 mm), and median target expansion diameter was 28 mm (range: 23-30 mm). Two procedural incidents occurred during stent delivery and were percutaneously treated. Stent stability was documented during TPVRs immediately performed in 14 patients. Median stent shortening was 13.7%, and median percentage of intended stent expansion was 95.9%. There was no stent fracture on the short-term follow-up (median: 4.5 months).
Conclusions:
We report the first implantations of Optimus-XXL stents in dysfunctional RVOTs with excellent preliminary results. Optimus-XXL should be considered as a valuable adjunct in the armamentarium for routine and complex TPVR procedures.