Purpose:
To evaluate the reliability of spectral-domain optical coherence tomography (SD-OCT; RTVue XR; Optovue, Inc., Fremont, CA, USA) for thickness mapping of the entire cornea (CT), corneal epithelium (ET). and corneal stroma (ST) over a 9-mm zone in healthy eyes. We sought to develop reference values for different age groups and elucidate potential sex- and age-dependent characteristics of corneal sublayer pachymetry maps.
Methods:
Three consecutive SD-OCT scans were obtained in 166 healthy right eyes (mean age = 50 ± 20 years). The thickness maps contain 25 sectors over a 9-mm diameter zone. To test measurement reliability, intraclass correlation coefficients (ICC), coefficients of variation (CoV), and within-subject standard deviations (WSSD) were calculated.
Results:
CT, ET, and ST ICCs ranged from 0.961 to 0.998, 0.896 to 0.945, and 0.955 to 0.998, respectively. CoV values for CT, ET, and ST ranged between 0.3 and 1.5%, 1.6 and 4.2%, and 0.4 and 1.7%, respectively. WSSD ranged from 6 to 41, 4 to 8, and 7 to 46 µm, respectively. A negative correlation was found between age and ET (p < 0.05) but not between age and ST or CT. No gender-related differences in CT, ET, or ST were detected. CoV of CT, ET, and ST measurements showed a positive correlation with age in 28, 64, and 28% of the sectors, respectively.
Conclusion: